AEMC MN114 AC Current Probe Model MN114 (10A, 100mV/A, Lead)
The MN114 is a general purpose mini probe which offers 1mA to 10AAC, continuous cycle current range. The output signal is 100mVAC/AAC (1V at 10A). It was designed to optimize measurements on 5A signals. The output termination is a 5ft lead with safety 4mm banana plug.
Applications:
| Operating Range | |
| Current Range | 1m/A to 10AAC, continuous duty cycle |
| Frequency Range | 30Hz to 5kHz |
| Common Mode Voltage | 600VAC |
| Current Overload | 20A continuous |
| Load Impedance | 100kW min |
| Reference Conditions | 23°C ± 3°K, 20 to 85% RH, external magnetic field <40A/m, no DC component, no external current carrying conductor, test sample centered. Load impedance 1mW |
| Working Voltage | 600VAC |
| Output Signal | |
| Output Signal | 100mVAC/AAC (1V at 10A) |
| Accuracy | |
| Range Value | 1mA to 10A / 40Hz to 3kHz |
| Accuracy % of Reading | 2% Reading ± 2mA / 3.5% Reading ± 2mA |
| Phase | |
| Range Value | 45 to 65Hz |
| Current Range Value | 10mA to 10A |
| Phase Shift | £8° |
| Mechanical Specifications | |
| Colors | Dark gray handles with red cover |
| Dimensions | 1.26 x 4.5 x 0.87" (32 x 115 x 22mm) |
| Material | Handles: 10% Fiberglass charged polycarbonate UL94 V0 |
| Maximum Cable Diameter | 0.47" & Oslash; max. (12mm) |
| Operating Temperature | 14° to 122°F (-10° to 50°C) |
| Output | 5 ft (1.5m) lead with safety 4mm banana plug |
| Storage Temperature | -40° to 176°F (-40° to 80°C) |
| Vibration | Test per IEC 68-2-6 |
| Weight | 5.6 oz (160g) |
| External Influences | |
| Influence of Temperature | <0.2% per10K° |
| Safety Reference Specifications | |
| Safety Ratings | 600V working voltage; 600V max common mode between output and ground; 3kV 50/60Hz dielectric for 1 min |